4 gain test, Gain test – Metrohm NIRS XDS MasterLab Analyzer User Manual
Page 101
![background image](/manuals/587149/101/background.png)
▪▪▪▪▪▪▪
99
The four standards are plotted out in the USP
format, showing repeatable response for
each standard. The x-axis, A REF, is from the
initial observation (during IPV Setup). The y-
axis, A OBS, is the current observation.
The slope must be 1.0, +/- 0.05. The zero
intercept must be 0.0 +/_0.05. The box
shows the results for this observation.
Tabs allow the user to check at two
wavelengths: 1200 and 1600nm.
Click on the Photometric Linearity Test
Results tab to return to the tabular
information.
The user should be aware that USP tolerances for Photometric Linearity are quite relaxed, compared
to FOSS NIRSystems specifications for IPV testing. Therefore, there may be times when the USP test
passes, and IPV does not meet FOSS NIRSystems specifications. This may be due to instrument
changes such as dirty windows, failing lamp, or placement errors. In such cases, Vision advises that
Reference Standardization may be required.
7.2.4
Gain Test
The Gain Adjust feature can be a useful diagnostic tool, though it is not required for normal
operation. Technically, gain is never adjusted on the XDS MasterLab Analyzer. The name of the test
comes from an adjustment capability that was required with older systems. With XDS, this program
reports gain information for the NIR and visible regions. No adjustment is performed.
To start Gain Adjust, click on Diagnostics,
then Gain Adjust.
The instrument must be connected and in
communication for this to function.