beautypg.com

12 servicing the detector, 13 rinsing the sample path – Metrohm 930 Compact IC Flex ChS/Deg User Manual

Page 89

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

930 Compact IC Flex ChS/Deg (2.930.1260)

■■■■■■■■

81

5.12

Servicing the detector

Follow the maintenance instructions in the the detector's manual.

5.13

Rinsing the sample path

Before a new sample is measured, the sample path has to be rinsed with it
long enough to prevent the measuring result from being falsified by the
previous sample (sample carry-over).

The time required to rinse the sample path with the new sample is called
the rinsing time. The rinsing time depends on the transfer time.

The transfer time corresponds to the time the sample needs to flow from
the sample vessel to the end of the sample loop. The transfer time
depends on the following factors:

The pump capacity of the peristaltic pump or Dosino

The total capillary volume

The volume of the gas removed from the sample by the sample degas-
ser

The transfer time can be determined as follows:

Determining the transfer time

1 Emptying the sample path

Pump air through the sample path (pump tubing, tubing connec-
tions, capillary in the degasser, sample loop) for several minutes until
all liquid is displaced by the air.

2 Aspirating the sample and measuring the time

Aspirate a sample typical for the later application and use a stop-
watch to measure the time the sample needs to go from the sample
vessel to the end of the sample loop.

The time upon stopping the watch corresponds to the "transfer
time".

If the sample is injected automatically, the rinsing time has to be at least
three times the transfer time.