3.10 Injection valve

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

930 Compact IC Flex ChS/Deg (2.930.1260)

Figure 12

Exchanging the sample loop

1

Pressure screw
Fastened to Port 6.

2

Sample loop

3

Pressure screw
Fastened to Port 3.

Exchanging the sample loop

1 Removing the existing sample loop

Unscrew the pressure screws (6.2744.010) at Port 3 and Port 6.

Remove the sample loop.

2 Installing a new sample loop

Fasten one end of the sample loop to Port 3 using a PEEK pres-
sure screw (6.2744.010).

Use the second PEEK pressure screw (6.2744.010) to fasten the
other end of the sample loop to Port 6.