5.11 Suppressor
■■■■■■■■■■■■■■■■■■■■■■
76
■■■■■■■■
930 Compact IC Flex ChS/Deg (2.930.1260)
■
In the software, use the Step command to switch to the next
suppressor unit and regenerate it as described above.
■
As soon as all three suppressor units have been regenerated, dis-
connect the capillary labeled regenerant from the coupling.
3 Rinsing the suppressor
After regeneration, each of the three suppressor units must be rinsed
with degassed ultrapure water for about 15 minutes.
■
Connect the capillary labeled rinsing solution to the outlet of
the high-pressure pump using a coupling (6.2744.040).
■
Set the flow of the high-pressure pump to 0.5 mL/min in the soft-
ware.
■
Switch on the high-pressure pump.
If the pressure decreases during rinsing, slowly increase the flow
of the pump to a maximum of 2 mL/min. When doing so, make
sure that the pressure does not exceed 2 MPa!
■
After approx. 15 minutes, switch off the high-pressure pump.
■
In the software, use the Step command to switch to the next
suppressor unit and rinse it as described above.
■
As soon as all three suppressor units have been rinsed, disconnect
the capillary labeled rinsing solution from the coupling.
4 Connecting the suppressor to the IC system
■
Reconnect the capillaries of the suppressor labeled regenerant
and rinsing solution to the IC system.
■
Reconnect the outlet of the high-pressure pump to the IC system.
5.11.3.3
Cleaning the suppressor
In the following cases, it may be necessary to clean the suppressor:
■
Increased backpressure at the suppressor's connection tubing.
■
Blockage of the suppressor which cannot be eliminated (solutions can
no longer be pumped through the suppressor).
■
Jamming of the suppressor which cannot be eliminated (suppressor
can no longer be switched over).
Cleaning the suppressor
1 Disconnecting the suppressor from the IC system
■
Switch off the instrument.
■
Disconnect all capillaries of the suppressor from the IC system.