beautypg.com

Metrohm 930 Compact IC Flex ChS/Deg User Manual

Page 87

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

930 Compact IC Flex ChS/Deg (2.930.1260)

■■■■■■■■

79

Replacing parts of the suppressor

1 Disconnecting the suppressor from the IC system

Switch off the instrument.

Disconnect all capillaries of the suppressor from the IC system.

2 Dismantling the suppressor

Unscrew union nut (21-1) from the housing (21-4).

Pull the connecting piece (21-2) out of the housing together with
the the rotor (21-3).
If rotor A gets stuck in the housing, you can knock it loose as fol-
lows:
Put a pointed object into the slot in the housing and use it to
knock rotor A loose.

Detach the connecting piece from the rotor.

3 Cleaning the new rotor

Clean the surface of the new rotor (21-3) with ethanol using a
lint-free cloth.

4 Inserting the new rotor

CAUTION

Rotor A may be destroyed during start-up if not inserted correctly.

Insert the new rotor (21-3) into the housing (21-4) in such a way
that the tubing connections on the rear of rotor fit into the corre-
sponding recesses inside the housing and one of the three holes
of the rotor is visible from below in the slot of the housing (21-5).