5 Operation and maintenance
930 Compact IC Flex ChS/Deg (2.930.1260)
■■■■■■■■
61
Spare parts
If the inlet valve cannot be cleaned, you will need a new inlet valve
(6.2824.170) for this step.
Cleaning the inlet valve
1 Removing the inlet valve
■
Start by loosening the inlet valve holder with the adjustable
wrench (1) and then unscrew it by hand (2) and remove it.
■
Remove the inlet valve from the inlet valve holder (3).
2 Cleaning the inlet valve
■
Start by rinsing the inlet valve in the direction of eluent flow using
a wash bottle filled with ultrapure water, RBS™ solution or ace-
tone. (The direction of eluent flow is marked on the valve by an
arrow.)
The rinsing solution must come out at the valve outlet.
If no more solution comes out, then the valve is blocked.
■
Rinse the inlet valve in the direction opposite the eluent flow
using a wash bottle filled with ultrapure water, RBS™ solution or
acetone.
The rinsing solution should only come out at the valve outlet.
■
The rinsing effect is further increased through a short treatment in
an ultrasonic bath.
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)