beautypg.com

2 connecting the leak sensor, 5 connecting the eluent bottle, Connecting the leak sensor – Metrohm 930 Compact IC Flex ChS/Deg User Manual

Page 28: Connecting the eluent bottle

background image

3.5 Connecting the eluent bottle

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

930 Compact IC Flex ChS/Deg (2.930.1260)

Route the loose end into a waste container.

3.4.2

Connecting the leak sensor

Plugging in the leak sensor connection cable

The leak sensor connection cable is coiled up in the base tray.

1 Pull the leak sensor connection cable out of the base tray as far as

needed.

2 Plug the plug for the leak sensor connection cable into the leak sen-

sor connection socket (labeled Leak Sensor).

3.5

Connecting the eluent bottle

The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing . The eluent aspiration tubing is installed on the input for the elu-
ent degasser.

The tubing has to be be fed out of the instrument through a suitable
opening (see "Openings on the back panel", page 12) before the loose
end can be connected to the eluent bottle.

Accessories

For this step you need the following accessories:

These parts are part of the accessory kit Vario/Flex ONE (6.5000.010).

Eluent bottle (6.1608.070)