2.4 Connecting to a modular gradient IC system
828 IC Dual Suppressor/ 8.828.1003 Instructions for Use
15
2.4.2 Connecting the IC Dual Suppressor
The following diagram shows how the system is arranged:
Figure 6:
Flow diagram of the MIC 10
All the components should be connected as shown in Figure 6, which
also provides a flow diagram for the sample.
The installation of the other components is described in detail in the
appropriate Instructions for Use.
The 828 IC Dual Suppressor is inserted between the separation column
and the detector. The following connections must be made:
Please ensure that the liquid connections between the separation
column outlet, the Dual Suppressor and the detector inlet are as short
as possible and use a narrow diameter capillary, e.g. 6.1831.010
PEEK Capillary (included as standard). In this way you minimize the
dead volumes and reduce the peak broadening of the signal.
IC Pumps
IC Detector
IC Dual Suppressor
Sample
Gradient
mixing capillary
Pulsation
dampener
T-connector
Separation
column
Sample
loop
A Trap column
Valve
Inline filter