beautypg.com

20 connecting and rinsing the guard column – Metrohm 930 Compact IC Flex Oven/SeS/PP/Deg User Manual

Page 61

background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)

■■■■■■■■

51

5 Setting the contact pressure of the peristaltic pump

Set the contact pressure of the peristaltic pump (see "Setting the
contact pressure correctly", page 40)
.

6 Rinsing the instrument without columns

Rinse the instrument (without columns) with eluent for 5 minutes.

3.20

Connecting and rinsing the guard column

Guard columns protect the separation column and substantially increase
its service life. Guard columns available from Metrohm are either actual
guard columns or what are known as guard column cartridges that can be
used together with a cartridge holder. The process for installing a guard
column cartridge in the associated holder is described in the leaflet for the
guard column.

NOTE

Metrohm recommends working with guard columns at all times. They
protect the separation column and can be replaced regularly as needed.

NOTE

Information regarding which guard column is suitable for your separa-
tion column can be found in the Metrohm IC Column Program
(which is available from your Metrohm representative), the leaflet provi-
ded along with your separation column, the product information about
the separation column at

http://www.metrohm.com

(Ion Chromatogra-

phy product area), or obtained directly from your representative.

CAUTION

New guard columns are filled with solution and sealed with stoppers or
caps on both sides.

Before inserting the guard column, ensure that this solution can be
mixed with the eluent being used (follow information from the manu-
facturer).