Metrohm 930 Compact IC Flex Oven/SeS/PP/Deg User Manual
Page 111

■■■■■■■■■■■■■■■■■■■■■■
6 Troubleshooting
930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)
■■■■■■■■
101
Problem
Cause
Remedy
The eluent contains gas
bubbles.
■
Inspect the connectors for the eluent
degasser .
■
Deaerate the high-pressure pump (see
Chapter 3.19, page 49).
Baseline increasing
strongly.
Suppressor – Reduced
capacity.
Regenerate the suppressor (see Chapter
5.12.3.2, page 85).
SPM – Reduced capacity.
Regenerate the SPM (see Chapter 5.12.3.2,
page 85).
Chromatograms
have poor resolution
Separation column –
Diminished separating effi-
ciency.
■
Regenerate the separation column (see
Chapter 5.17.4, page 96).
■
Replace the separation column (see "Con-
necting the separation column", page 55).
Extreme spread of
the peaks in the
chromatogram.
Splitting (dual
peaks)
Capillary connections –
dead volume in the system.
Check the capillary connections (see Chapter
3.2, page 13) (use PEEK capillaries with an
inner diameter of 0.25 mm between the injec-
tion valve and detector).
Guard column – Dimin-
ished efficiency.
■
Replace the guard column (see Chapter
3.20, page 51).
Separation column – Dead
volume at the column
head.
■
Install the separation column in the oppo-
site flow direction (if permitted by the leaf-
let) and rinse into a beaker.
■
Replace the separation column (see "Con-
necting the separation column", page 55).
Precision problems -
the measured values
are highly scattered.
Injection valve – Sample
loop.
Inspect the installation of the sample loop (see
"Optional: Exchanging the sample loop",
page 26).
Sample – There are gas
bubbles in the sample.
Use the sample degasser .
Sample – The rinsing vol-
ume is too small.
Increase the rinsing time (see Chapter 5.16,
page 94).
Injection valve – Defective.
Request Metrohm Service.
MCS – The vacuum is too
low.
■
Check the connectors. If they are OK:
■
Contact Metrohm Service.