beautypg.com

1 introduction, 1 instrument description, Instrument description – Metrohm 930 Compact IC Flex Oven/SeS/PP/Deg User Manual

Page 11

background image

■■■■■■■■■■■■■■■■■■■■■■

1 Introduction

930 Compact IC Flex Oven/SeS/PP/Deg (2.930.2560)

■■■■■■■■

1

1 Introduction

1.1

Instrument description

The 930 Compact IC Flex is a professional ion chromatograph. It is distin-
guished by:

Its intelligence: All of the functions are monitored, optimized and
documented in an FDA-compatible manner. Intelligent components,
such as iColumns, save important data onto a chip.

Its compact design: It has a small footprint.

Its transparency: All components are easily accessible and located for
simple visibility and can be monitored during operation through a large
window.

Its safety: The wet end and electronics are physically separated, thereby
preventing fluids from coming into contact with the electronics to a
large extent. A leak sensor is integrated into the wet end.

Its environmental compatibility.

Its low noise emissions.

The intelligent MagIC Net™ software

The 930 Compact IC Flex is operated solely using the MagIC Net software.
A USB cable is used to connect the instrument to a computer with
MagIC Net installed. The intelligent software detects the instrument auto-
matically and checks its functionality. The software controls and monitors
the instrument, evaluates the measured data and manages it in a data-
base.

The 930 Compact IC Flex Oven/SeS/PP/Deg consists of the following mod-
ules:

Leak sensor
The leak sensor detects leaking liquid that collects in the instrument's base
tray. Liquid that leaks in the instrument is routed to the base tray using
drainage tubing and detected there.

Column oven
The column oven tempers the separation column and the eluent and
thereby provides stable measuring conditions. A column holder with a
chip reader is located in the column oven.

Eluent degasser
The eluent degasser removes gas bubbles and dissolved gases from the
eluent.