5 Technical data

757 VA Computrace – Hardware

60

VA Computrace Interface

Designation

6.2155.000 VA Computrace Interface

Type

Interface USB – VA Computrace Stand

Plug to 757

D-Sub with 37 pins

Mains connection

Voltage

100...240 V

Frequency

50...60 Hz

Power consumption

26 W

Fuse

2 × 1.0 ATH (to be replaced by Metrohm Service only

using the same type).

Additional electronic overload protection.

Safety specifications

Construction/testing

According to IEC 61010 / EN 61010 / UL 3101-1,

protection class 1

Safety directions

The Instructions for Use include information and

warnings which must be heeded by the user to

assure safe operation of the instrument.

Electromagnetic compatibility (EMC)

Emitted interference

Standards met:

EN55022 (class B), EN50081-1/2

Immunity to interference Standards met:

IEC61000-4-2 (level 4), IEC61000-4-3 (level 2),

IEC61000-4-4 (level 4), IEC61000-4-5 (level 2/3),

IEC61000-4-6 (level 3), IEC61000-4-8,

IEC61000-4-11, ENV50204, EN50082-1/2

Ambient temperature

Nominal operating range 0…+45 ° C
Storage, transport

–40…+70 ° C

Housing

Material of cover

Polyurethane rigid foam (PUR) with fire protection for

fire class UL94VO, FCH-free

Material of base

Steel, enamelled

Material of measuring

head arm

Steel, enamelled

Dimensions

Width

259 mm

Height

241 mm (417 mm with cover raised)

Depth

526 mm

Weight

9.7 kg (excl. accessories)