Metrohm 757 VA Computrace User Manual

Page 51

background image

3.9 Connection of the 813 Compact Autosampler

757 VA Computrace – Hardware

47

3 Connect 772 siphoning pump

• Cut 6.1826.100 pump tubing to a length of ca. 17 cm.
• Attach two 6.1820.050 tubing connectors to both ends of the 6.1826.100

pump tubing and install it at the first 772 Pump Unit (see Instructions for

Use 772).

• Screw a 6.1820.020 thread onto both 6.1820.050 connectors.
• Using a 6.1805.530 FEP tubing, connect the lower end of the pump

tubing on the siphoning pump to the 6.1808.000 connector at the 757 VA

Computrace Stand, to which the FEP tubing 34 is connected (see Fig. 16

and Fig. 17).

4 Connect 772 rinsing pump

• Cut 6.1826.100 pump tubing to a length of ca. 17 cm.
• Attach two 6.1820.050 tubing connectors to both ends of the 6.1826.100

pump tubing and install it at the second 772 Pump Unit (see Instructions

for Use 772).

• Screw a 6.1820.020 thread onto both 6.1820.050 connectors.
• Using a 6.1805.530 FEP tubing, connect the upper end of the pump

tubing on the rinsing pump to the 6.1808.000 connector at the 757 VA

Computrace Stand, to which the FEP tubing 119 is connected (see Fig.

16 and Fig. 17).

5 Connect waste container

• Unscrew red filling connection from first 6.1621.000 container.
• Screw 6.1618.050 threaded adapter to the container.
• Screw 6.1602.105 siphon onto 6.1618.050 threaded adapter.
• Cut 6.1819.020 FEP tube to a length of ca. 10 cm and insert from above

into the smallest opening of the 6.1602.105 siphon.

• Screw a 6.1805.530 FEP tubing into this opening of the 6.1602.105

siphon.

• Attach the other end of the 6.1805.530 FEP tubing at the upper end of the

pump tubing of the siphoning pump (see Fig. 16).

6 Connect storage container

• Unscrew red filling connection from second 6.1621.000 container.
• Using a funnel, add max. 10 L rinsing solution (normally ultrapure water

acidified with 100 µL conc. HCl/L) to the storage container through the

small vent opening.

• Screw 6.1618.050 threaded adapter to the container.
• Screw 6.1602.105 siphon onto 6.1618.050 threaded adapter.
• Remove screw nipple from 6.1829.020 FEP tube and insert it from above

into the smallest opening of the 6.1602.105 siphon.

• Screw a 6.1805.530 FEP tubing into this opening of the 6.1602.105

siphon.

• Attach the other end of the 6.1805.530 FEP tubing at the lower end of the

pump tubing of the rinsing pump (see Fig. 16).