beautypg.com

5 technical data – Metrohm 757 VA Computrace User Manual

Page 61

background image

5 Technical data

757 VA Computrace – Hardware

57

5 Technical data

Subject to changes !
The listed technical data apply to an ambient tempera-

ture of 25°C.

Brief characterization

PC-controlled system for voltammetry, set chemical workplace with potentiostat

and measuring amplifier.
With multi-mode electrode, rotating disk electrode (RDE) as option.
Tilt-back measuring arm, integrated drip pan.

Current measurement techniques

DC

Direct Current

NP

Normal Pulse

DP

Differential Pulse

SQW

Square Wave (10 ... 2000 Hz)

AC1

Phase-sensitive Alternating Current 1

st

harmonic

(1 … 250 Hz)

AC2

Phase-sensitive Alternating Current 2

nd

harmonic

(1 … 250 Hz)

PSA

Potentiometric Stripping Analysis (chronopotentiometry)

CV

Cyclic Voltammetry (digital ramp)

Potentiostat

Output voltage (AE)

± 12 V

Output current (AE)

± 35 mA

Sweep voltage range

± 5 V

Voltage resolution

150 µV

Input impedance(RE)

R ≥ 1·10

10

Input Bias Current (RE)

± 10 pA

Noise

typ. 200 pA
Circuit for measurement of the noise:

100 nF

AE RE

WE

Measurement mode: Differential Pulse

(method used: DPNoise.mth)