6 dosing device control, Dosing functions, 7 display interface signals – Metrohm 838 Advanced Sample Processor Manual User Manual
Page 59: Display remote interface signals, Display rs232 interface signals, Dosing device control, Display interface signals

2.7 Manual operation
Metrohm 838 Advanced Sample Processor, Operation
51
With the STIR command all of the connected stirrers can be con-
trolled. The first parameter selects the stirrer. In addition to switching
on and off, the stirring time can be chosen (in seconds).
2.7.6 Dosing device control
DOS
6
manual operation
DOS: 1.* : : 1 mL
1.1…3.4,
1.*…3.*, *.*
fill,release, prep
empty, eject, endVol,
compen., port,
Dosing functions
The DOS command is used for controlling Dosimats and Dosinos.
The 1st parameter selects the dosing device and the corresponding
dosing port, at which the required function is to be carried out. In
addition to the dosing volume, several dosing functions can be cho-
sen. For details, see section 2.4.3.
2.7.7 Display interface signals
1
SCAN
manual operation
SCN:Rm :00000000
Rm,RS 0,1
Display remote interface
signals
The
SCN:Rm
command causes the signal states of the incoming re-
mote lines to be displayed in binary format (
1
= line active,
0
= line
inactive).
manual operation
SCN:RS
Rm,RS ASCII
characters
Display RS232 interface sig-
nals
The
SCN:RS
command causes the character string received at the
RS232 interface to be displayed line by line (14 characters).