2.1 Operating principles

10

Metrohm 838 Advanced Sample Processor, Operation

STOP

The key ends a method.

If a sample series is stopped manually with then the final

sequence of a method will not be carried out. When the key

is pressed the functions listed in the parameter menu under

>manual

stop

will be carried out.

HOLD

LEARN

The key interrupts a method sequence.

Connected peripheral devices will not be stopped automatically. Only

the method sequence will be interrupted. In the HOLD condition a

method can be completely terminated with or continued

with .

After an error message in the method sequence the Advanced Sam-

ple Processor switches automatically to the HOLD condition after

.