2.2
Configuration
Metrohm 838 Advanced Sample Processor, Operation
21
>recall rack
beaker sensor:
off
Selection of the beaker sensor
Tower, SwingH, off
Not implemented.
>recall rack
rack offset
0.00°
Correction for the rotation angle of the
sample rack *
)
-5.00…0.00…5.00°
*
)
All the above lift positions and the rack offset can be set by using the
Special beaker positions
>>recall rack
>>>special positions
Submenu special beaker positions
Open with
Special beaker positions are predefined places on a sample rack that are not
treated as sample positions. They can be occupied by rinsing beakers or
conditioning beakers and selectively addressed during a run sequence. Up
to 16 special beaker positions can be defined per rack. For each special
beaker the work position of the lift and the beaker radius can be set, see
above.
Selection of special beaker
>>>special positions
special beaker
1
1…16
Rack position of the special
beaker
>>special positions
1
rack position
0
etc. up to special beaker 16
0…number Pos.
0 = not defined
>rack definitions
>>store rack
Save rack definitions
In order to store modifications to the definition of a sample rack the submenu
>>store rack
is selected.
>store rack
name: 6.2041.310
Name of rack
10 ASCII characters
With the key you can select one of the existing rack names. The
rack data is stored with
of any rack name can be made directly via the numerical keys or in the text