beautypg.com

2 configuration, 1 miscellaneous, Configuration – Metrohm 838 Advanced Sample Processor Manual User Manual

Page 24: Miscellaneous

background image

2.2 Configuration

16

Metrohm 838 Advanced Sample Processor, Operation

2.2 Configuration

CONFIG

Main

menu:

configuration

>auxiliaries

• Open submenu with

configuration

>tower 1

• Use < > or < > to move up and down by one

menu item

configuration

>rack definitions

• Use or to move to the first or last

menu item

configuration

>Dosing units

• Use to return to operating condition

configuration

>RS232 settings

2.2.1 Miscellaneous

configuration

>auxiliaries

Submenu for basic settings

Open the submenu with

>auxiliaries

dialog: english

Select dialog language

english, deutsch,

français,

español


>

auxiliaries

display contrast

3

Set display contrast

0…3…7

0 = low contrast

7 = high contrast

>auxiliaries

beeper: ON

Beep for warnings on or off

ON,OFF

>

auxiliaries

external Start:

off

Switch on external start via Remote line

(Input 7)

on, off

Normally the Sample Processor takes over the complete run control for a

sample series as the "Master". This may be unwanted, particularly if an

automated system has been set up that includes instruments that are not

from Metrohm. This is why the Sample Processor can also be started from

any external instrument via the Remote interface.

With

external Start: ON

the run sequences of the Sample Processor will

be started as soon as input line 7 is activated at the Remote interface (low

level).

Use

for next higher

level