7 manual operation, Position vessel / turn rack, 2 moving the lift – Metrohm 838 Advanced Sample Processor Manual User Manual
Page 57: Manual operation, Turning the sample rack / positioning the samples, Moving the lift

2.7 Manual operation
Metrohm 838 Advanced Sample Processor, Operation
49
2.7 Manual operation
2.7.1 Turning the sample rack / Positioning the samples
NEXT
PREV
With the < NEXT> and < PREV> keys the sample rack
can be rotated forward or backward by one position.
< /NEXT> moves the next sample position (SAMPLE+1) to
the tower. < /PREV> moves the previous sample position
(SAMPLE-1) to the tower.
8
MOVE
Handbetrieb
MOVE 1 : Probe
sample,
ext.1…ext.4
spec.1…16
prev.,
next
+swing, -swing
+rotate, -rotate
1…999
+/-1…999
Position vessel /
Turn rack
With the MOVE command rack positions are moved to, i.e. the
turntable rotates until the sample rack is positioned so that the
selected rack position is placed under the needle. With LECT> the numerical rack position as well as the predefined current sample (SAMPLE command) or the special beakers 1- 16 can be chosen. Furthermore, with the robotic arm external position (not on the rack) can be accessed. More details about the MOVE command you can find in section 2.4.1. For security reasons turning the sample rack is only possible when the lift is in or above the shift position. The lift is auto- matically raised, if its current position is lower than the prede- fined shift position. 2.7.2 Moving the lift The keys < > and < > allow upward and downward move- ment of the lift. The lowest possible lift position is defined by the configuration parameter ' max. stroke path ' . HOME END The the upper limit.