beautypg.com

19 getinterfaceerror, Getinterfaceerror – Metrohm 846 Driver Toolbox User Manual

Page 22

background image

2.2 Functions

18

Dosing Interface USB Toolbox, DLL Reference

InRate

float

Filling rate [0.01…166 mL/min]

OutPort

long

Output port 1 [1…4]

OutVolume

float

Dosing volume 1 [0…20000 µL]

OutRate

float

Dosing rate 1 [0.01…166 mL/min]

SpecPort

long

Output Port 2 [1…4]

Specvolume

float

Dosing volume 2 [0…20000 µL]

SpecRate

float

Dosing rate 2 [0.01…166 mL/min]


Output parameters

eReturnstate

return

value

2.2.19 GetInterfaceError

GetInterfaceError(long IfNo, long MsbNo, long ErrorNumber,

char* ErrorCode, long ErrorBufSize)

Reads casual error messages which occur during Dosino actions.

Input parameters

IfNo

long

Dosing Interface [1… ]

MsbNo

long

Dosino at MSB[1…4]

Number

long

Error number [0…9]

ErrorBufSize

long

Number of characters


Output parameters

eReturnstate

return

value

ErrorCode

char*

Error code, format [GGG-CCC-K-III],
G = group, C = code, K = class,
I = index (1...4 or 255), see 2.3