beautypg.com

16 doscontinue, 17 du_prep, 18 empty – Metrohm 846 Driver Toolbox User Manual

Page 21: Doscontinue, Du_prep, Empty

background image

2.2 Functions

Dosing Interface USB Toolbox, DLL Reference

17

2.2.16 DosContinue

DosContinue(long IfNo, long MsbNo)

Continue a halted Dosino action.

Input parameters

IfNo

long

Dosing Interface [1… ]

MsbNo

long

Dosino at MSB[1…4]


Output parameters

eReturnstate

return

value

2.2.17 DU_Prep

DU_Prep(long IfNo, long MsbNo, long InPort, float InVolume,

float InRate, long OutPort, float OutVolume, float OutRate,

long SpecPort, float SpecVolume, float SpecRate)

Prepares a Dosing Unit for further use. A "Prep" cycle includes emptying the
dosing cylinder and rinsing and filling the tubings in one automated process.

Input parameters

IfNo

long

Dosing Interface [1… ]

MsbNo

long

Dosino at MSB[1…4]

InPort

long

Filling port [1…4]

InVolume

float

Fill tube volume [0…20000 µL]

InRate

float

Filling rate [0.01…166 mL/min]

OutPort

long

Output port 1 [1…4]

OutVolume

float

Dosing volume 1 [0…20000 µL]

OutRate

float

Dosing rate 1 [0.01…166 mL/min]

SpecPort

long

Output Port 2 [1…4]

Specvolume

float

Dosing volume 2 [0…20000 µL]

SpecRate

float

Dosing rate 2 [0.01…166 mL/min]


Output parameters

eReturnstate

return

value

2.2.18 Empty

Empty(long IfNo, long MsbNo, long InPort, float InVolume, float

InRate, long OutPort, float OutVolume, float OutRate, long

SpecPort, float SpecVolume, float SpecRate)

Empties the Buret Unit. Dosing cylinder and tubings are emptied in one auto-
mated process.

Input parameters

IfNo

long

Dosing Interface [1… ]

MsbNo

long

Dosino at MSB[1…4]

InPort

long

Filling port [1…4]

InVolume

float

Fill tube volume [0…20000 µL]