2.2 Functions
Dosing Interface USB Toolbox, DLL Reference
17
2.2.16 DosContinue
DosContinue(long IfNo, long MsbNo)
Continue a halted Dosino action.
Input parameters
IfNo
long
Dosing Interface [1… ]
MsbNo
long
Dosino at MSB[1…4]
Output parameters
eReturnstate
return
value
2.2.17 DU_Prep
DU_Prep(long IfNo, long MsbNo, long InPort, float InVolume,
float InRate, long OutPort, float OutVolume, float OutRate,
long SpecPort, float SpecVolume, float SpecRate)
Prepares a Dosing Unit for further use. A "Prep" cycle includes emptying the
dosing cylinder and rinsing and filling the tubings in one automated process.
Input parameters
IfNo
long
Dosing Interface [1… ]
MsbNo
long
Dosino at MSB[1…4]
InPort
long
Filling port [1…4]
InVolume
float
Fill tube volume [0…20000 µL]
InRate
float
Filling rate [0.01…166 mL/min]
OutPort
long
Output port 1 [1…4]
OutVolume
float
Dosing volume 1 [0…20000 µL]
OutRate
float
Dosing rate 1 [0.01…166 mL/min]
SpecPort
long
Output Port 2 [1…4]
Specvolume
float
Dosing volume 2 [0…20000 µL]
SpecRate
float
Dosing rate 2 [0.01…166 mL/min]
Output parameters
eReturnstate
return
value
2.2.18 Empty
Empty(long IfNo, long MsbNo, long InPort, float InVolume, float
InRate, long OutPort, float OutVolume, float OutRate, long
SpecPort, float SpecVolume, float SpecRate)
Empties the Buret Unit. Dosing cylinder and tubings are emptied in one auto-
mated process.
Input parameters
IfNo
long
Dosing Interface [1… ]
MsbNo
long
Dosino at MSB[1…4]
InPort
long
Filling port [1…4]
InVolume
float
Fill tube volume [0…20000 µL]