1.3 Metrohm 700/800 Dosinos and Dosing Units

8

Dosing Interface USB Toolbox, Overview

1.3.4 Standard occupancy of the Dosino ports:

Port 1

Dosing outlet; M6 threaded connection on the left-hand side of
the housing.

The liquid is ejected through a dosing tip or a titration tip.

Port 2

Filling inlet; M6 threaded connection on the base of the dosing
unit.

The liquid is aspirated from a storage container.

Port 3

Not assigned; M6 threaded connection on the right-hand side of
the housing.

Port 4

Special functions; narrow connection nipple on the base of the
dosing unit. This can be used during the 'PREP' function for
ejecting the liquid. When the dosing unit is emptied port 4 is used
as the air inlet.

VENT

Vents the storage container; M6 threaded connection at the front.
A drying tube can be connected to it, e.g. filled with molecular
sieve or soda lime.