beautypg.com

Figure 43, Dismantling valve – Metrohm 850 Professional IC AnCat MCS User Manual

Page 90

background image

4.5 High pressure pump

■■■■■■■■■■■■■■■■■■■■■■

80

■■■■■■■■

850 Professional IC – AnCat – MCS

Place the valve with the seal faced downwards above the recess
in the holder.

Push the valve components out of the valve housing using the
needle of the tool.

1

3

4

2

Figure 43

Dismantling valve

1

Needle
For pushing the valve components out of
the valve housing.

2

Valve

3

Holder

4

Recess
For collecting the valve components.

The components of the valve are collected in the recess of the
holder.

Note

The components of the valve are very small. In order not to lose
them, put the components into a dish.

The inlet valve and the outlet valve consist of the same, just differ-
ently arranged components (see Figure 44, page 81).