beautypg.com

2 pump tubing connection with filter, Figure 47, Pump tubing connection – changing the filter – Metrohm 850 Professional IC AnCat MCS User Manual

Page 105

background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation and maintenance

850 Professional IC – AnCat – MCS

■■■■■■■■

95

Order
number

Name

Material

Inner
diameter

Use

6.1826.330

Pump tubing LFL
(orange/white), 3-
stopper

PVC (Tygon)

0.64 mm

No special applications.

6.1826.340

Pump tubing LFL
(black/black), 3-stop-
per

PVC (Tygon)

0.76 mm

For the sample solution in
inline dialysis.

6.1826.360

Pump tubing LFL
(white/white), 3-stop-
per

PVC (Tygon)

1.02 mm

For sample transfer.

6.1826.380

Pump tubing LFL
(gray/gray), 3-stopper

PVC (Tygon)

1.25 mm

For inline sample dilution.

6.1826.390

Pump tubing LFL (yel-
low/yellow), 3-stop-
per

PVC (Tygon)

1.37 mm

For the sample solution in
inline ultrafiltration.

4.12.2.2

Pump tubing connection with filter

The 6.2821.130 filters (47-2) should be changed every 3 months, more
frequently at higher backpressure.

1

2

3

Figure 47

Pump tubing connection – Changing the filter

1

Tubing olive

2

Filter 6.2821.130
Packaging contains 10 items.

3

Filter housing

Replacing the filter

1 Unscrewing filter screw

Screw the filter screw (47-3) out of the tubing olive (47-1) with
the aid of two 6.2621.000 adjustable wrenches.

2 Replacing the filter

Remove the old filter (47-2) with tweezers.

Place the new filter (47-2) flat in the tubing olive (47-1) with
tweezers.