2.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

850 Professional IC – AnCat – MCS

5 Connect the leak sensor (see Chapter 2.6.4, page 21).

6 Connect drainage tubings (see Chapter 2.6.5, page 22).

7 Connect the eluent bottle (see Chapter 2.8.1, page 27).

8 Install sample path connections.

Connect the sample degasser (if required) (see Chapter 2.13,
page 40)
.

Connect sample path connections to the injection valve (see
Chapter 2.14.2, page 42)
.

9 Installing the MSM (see Chapter 2.16, page 47) – with associated

peristaltic pump (see Chapter 2.17, page 50).

10 Connect the MCS (see Chapter 2.18.2, page 55).

11 Connect detector capillaries (see Chapter 2.19, page 58).

12 Mains connection (see Chapter 2.20, page 61).

13 Connect the instrument to the PC (see Chapter 2.20.1, page 61).

14 Initial start-up (see Chapter 3.1, page 66).

15 Install the guard column (if used) (see Chapter 2.21, page 62).

16 Install the separation column (see Chapter 2.22, page 64).