background image

4.13 Metrohm CO

2

suppressor (MCS)

■■■■■■■■■■■■■■■■■■■■■■

96

■■■■■■■■

850 Professional IC – AnCat – MCS

3 Mounting filter screw

Screw the filter screw (47-3) back into the tubing olive (47-1) and
tighten by hand. Then additionally tighten with two 6.2621.000
adjustable wrenches.

4.13

Metrohm CO

2

suppressor (MCS)

4.13.1

Replacing the CO

2

adsorption cartridge

The CO

2

adsorption cartridge 6.2837.000 (32-4) must be replaced regu-

larly. This is because of blockages and losses in capacity.

Blockage
Moisture can block the CO

2

adsorption cartridge. This is revealed by a

change in color of the cartridge material (the orange part becomes color-
less). As the air flow is reduced, the vacuum becomes lower. To protect
the CO

2

adsorption cartridge, an H

2

O adsorption cartridge (32-7) is instal-

led upstream. Regular regeneration (see Chapter 4.13.2, page 96) of
the H

2

O adsorption cartridge extends the service life of the CO

2

adsorp-

tion cartridge.

Capacity loss
The adsorption capacity of the CO

2

adsorption cartridge is limited.

Depending on the period of operation and laboratory environment, the
adsorption capacity will diminish over time. This is manifested in a rising
baseline (as more CO

2

reaches the detector).

4.13.2

Regenerating the H

2

O adsorption cartridge

The function of the H

2

O adsorption cartridge is to protect the CO

2

adsorp-

tion cartridge against moisture. The service life of the H

2

O adsorption car-

tridge depends on the moisture content of the ambient air. Moisture
reduces the capacity of the H

2

O adsorption cartridge (which can be

observed by a change in color). Before the color changes in the entire fill-
ing material (from orange to colorless, with Fluka Order No. 94098), the
H

2

O adsorption cartridge should be regenerated (see information sheet).

The filling material is replaced during regeneration:

1 Allow material to dry loose (not in cartridge) at 140 °C overnight and

refill. Or dispose of the old material, and fill with new material.

2 Cover the packed material with wadding.