Metrohm 850 Professional IC AnCat MCS User Manual
Page 112
5.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
102
■■■■■■■■
850 Professional IC – AnCat – MCS
Problem
Cause
Remedy
MCS – vacuum pump
defective.
Contact the Metrohm Service.
Background conduc-
tivity too high
MSM – not connected.
Connect the MSM (see Chapter 2.16, page
47).
MCS – not connected.
Connect the MCS.
Incorrect eluent.
Change eluent (see Chapter 4.4.2.3, page 72).
MSM – regeneration or
rinsing solution flow prob-
lems.
Check the flow of the regeneration and rinsing
solution (see Chapter 2.16.2, page 47).
Data of the separa-
tion column cannot
be read.
Column chip contamina-
ted.
Clean contact surfaces of the column chip
(with alcohol).
Column chip defective.
1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.
Individual peaks
greater than expec-
ted
Sample – Carry-over of the
samples from previous
measurement.
Rinse system longer between two samples.
MSM – No (or insuf-
ficient) pumping of
regeneration or rins-
ing solution
Leak in the system.
Check connections.
Peristaltic pump – contact
pressure too weak.
Correctly set contact pressure (see "Set flow
rate", page 54).
Peristaltic pump – filter
blocked (see Figure 29,
page 52).
Replace the filter (see "Replacing the filter",
page 95).
MSM – backpressure too
high.
Clean the MSM (see Chapter 4.11.3.2, page
89) or replace parts (see Chapter 4.11.3.3,
page 92).
Peristaltic pump – pump
tubing defective.
Replace pump tubing (see Figure 28, page 51).
Poor reproducibility
of the retention
times
Eluent – Leak in the eluent
path.
Check the eluent path.
Eluent – Blockage in the
eluent path.
Check the eluent path.
Chromatograms
have poor resolution
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 4.15.4, page 98).