2.8 Eluent

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

850 Professional IC – AnCat – MCS

Fasten the bottle attachment (13-2) on the eluent bottle (15-10).
The aspiration filter (15-6) must rest on the base of the eluent
bottle.

6 Mounting the adsorber tube

Note

In the case of alkaline eluents and eluents with lower buffer
capacity, the eluent bottle must be equipped with a CO

2

adsorber

(15-4).

First, place a piece of wadding (15-3), then the CO

2

adsorber

(15-4) in the large opening of the adsorber tube (15-2) and close
with the plastic cover.

Fasten the adsorber tube (15-2) using the SGJ clip (15-12) onto
the bottle attachment (15-11).