beautypg.com

0 diagnostics and troubleshooting, 1 incorrect data – Rockwell Automation 57C402 25-115V AC/DC Low Power Output Module User Manual

Page 17

background image

5Ć1

5.0 DIAGNOSTICS AND

TROUBLESHOOTING

This section explains how to troubleshoot the module and field

connections.

5.1

Incorrect Data

Problem: The device connected to the output is either always off,

always on, or actingdifferent than expected. The possible causes of

this are a module in the wrongslot, a programmingerror, or a

malfunctioningmodule. It is also possible that the output is either not

wired or wired to the wrongdevice. Use the followingprocedure to

isolate the problem:
Step 1.

Verify that the output module is in the correct slot and that

the I/O definitions are correct.
Refer to figure 3.2. Verify that the slot number being

referenced agrees with the slot number defined in the

configuration. Verify that the register number is 0. Verify

that the bit number refers to the proper bit. Note that the

bit number and the wire number are not the same.
For remote I/O installations, also verify that the master slot

and remote drop number are defined correctly.

Step 2.

Verify that the power supply is functional.
Confirm that all connections at the terminal strip are tight.

Connect a voltmeter to the proper points on the terminal

strip and toggle the device. The voltmeter should alternate

between 0 and 115 volts. If this does not happen, there is

a problem with either the external device, the DC power

supply, or the wiringto the terminal strip.
Connect the programming terminal to the system and run

the ReSource software.
Stop all tasks that may be running.
Confirm that all connections at the terminal strip are tight.

Connect a voltmeter to the power supply connections on

the terminal strip. If the voltage is not correct, there is a

problem with the power supply or the wiringto the

terminal strip.
Check the cable for continuity between the faceplate

connector and the terminal strip.

Step 3.

Verify that the module can be accessed.
Connect the programming terminal to the system and run

the ReSource Software.
Stop all tasks that may be running.
Use the I/O MONITOR function. Toggle the output device

and determine whether the bit is changing state by

observingthe condition of the LED on the module. If the

LED does not change state, the hardware is

malfunctioning.