Current source calibration – Rockwell Automation 1794-IRT8XT FLEX I/O Thermocouple/RTD /Millivolt Input Module User Manual User Manual
Page 59
![background image](/manuals/579278/59/background.png)
Publication 1794-6.5.12 - September 2011
Calibrate Your Module 47
2. Apply power to the module for 20 minutes before calibrating.
3. After the connections stabilize, initiate a write word with the appropriate
EDT command location (write word 3, bits 00…15) as shown above.
4. Monitor the value in the EDT response location (read word 11,
bits 00…15) for an echo of the EDT command. Depending on which
command you sent, it should show 242X Hex.
If the EDT response word reads 80FF (hex), repeat step 3. Make sure to
allow for sufficient time for the module to respond to your request. If
there is no change, calibration has failed. Check the wiring and try again.
5. Set the precision millivolt source to the value required for a gain of 2.
Repeat steps 3 and 4 for gain 2. Repeat for each gain setting.
When all gain calibrations are successful, proceed to perform Current Source
Calibration on page 47.
Current Source Calibration
The current sources can be calibrated one at a time or all at once. To calibrate
all the current sources, proceed as follows:
1. Connect a 383
Ω
, 0.01% resistor across (H, +) and (L, -) of each input
channel (8 resistors).
2. Apply power to the module for 20 minutes before calibrating.
EDT Commands for Gain Calibration
All Channels
Single Channel
EDT Command
EDT
Command
Dec. or Hex
Bits 8
…
15
EDT Command
Data
Dec. or Hex
Gain
Selected
Input (mV)
Hex
Decimal
Bits
4
…
7
Bits
0
…
3
1
-320.00 + 0.064 mV
2420
9248
06
0
…
7
0
2
320.00 + 0.032 mV
2421
9249
06
0
…
7
1
4
110 + 0.016 mV
2422
9250
06
0
…
7
2
8
70.00 + 0.008 mV
2423
9251
06
0
…
7
3
16
29.00 + 0.004 mV
2424
9252
06
0
…
7
4
32
19.00 + 0.002 mV
2425
9253
06
0
…
7
5