Note: To recalibrate from any screen, hold anywhere on the screen for 10 seconds. You will see a
small progress bar at the bottom of the screen. When the progress bar reaches 100 percent, the
calibration screen will open.

Step 3:

The "Touch Screen

Calibration" screen will then appear
(See

Figure 12

). Press and hold on the

center of the touch target, release when
the touch target begins to flash. Repeat
with the next two touch targets.

Step 4:

The screen uses the new

calibration configuration. You can test
the calibration before saving by
pressing anywhere on the screen to
ensure the touch target appears where
you press. After testing, press the
"SAVE" button. Press the "BACK"
button to return to the "System
Configuration" screen.

Figure 12 - Touch Screen Calibration

Step 5:

Set date and time by pressing

"SET SYSTEM CLOCK" in "System
Configuration." Press “SAVE” when
completed.

Figure 13 - Set System Date and Time