Sunnex CS2050D (Dual) User Manual
Page 12

INSTALLATION AND OPERATING MANUAL
11
Table 1 Guideline for ceiling style mount
(Refers to Figures 8, 9 & 10)
CEILING LIGHT MOUNTING INSTRUCTIONS
WITH STANDARD EXTENSION Z-ARM
TO ACHIEVE FLOOR
TO LIGHT DISTANCE
OF
“A”
IF DROP CEILING
IS
“B”
MOUNT BOTTOM OF
TRANSFORMER
BRACKET
FROM FLOOR AT
“C”
GAP BETWEEN
DROP CEILING
AND CEILING ARM
“D”
MINIMUM 2”
REQUIRED
75 INCHES
(1905 mm
)
8.0 FEET
(
2440 mm
)
99.8 INCHES
(2535 mm
)
6.5”
(165mm)
86 INCHES
(
2184 mm
)
9.0 FEET
(
2740 mm
)
109.8 INCHES
(
2789 mm
)
4.5”
(114mm)
______________________________________________________________________________
CEILING LIGHT MOUNTING INSTRUCTIONS
WITH EXTRA-HIGH EXTENSION Z-ARM
TO ACHIEVE
FLOOR TO LIGHT
DISTANCE OF
“A”
IF DROP CEILING
IS
“B”
MOUNT BOTTOM OF
TRANSFORMER
BRACKET
FROM FLOOR AT
“C”
GAP BETWEEN DROP
CEILING AND
CEILING ARM
“D”
MINIMUM 2”
REQUIRED
72 INCHES
(
1828 mm
)
10 FEET
(
2280 mm
)
123.8 INCHES
(
3145 mm
)
2.5”
(63mm)
85 INCHES
(
2032 mm
)
11 FEET
(
3350 mm
)
133.8 INCHES
(
3399 mm
)
4.5”
(114mm)