INSTALLATION AND OPERATING MANUAL

11

Table 1 Guideline for ceiling style mount

(Refers to Figures 8, 9 & 10)

CEILING LIGHT MOUNTING INSTRUCTIONS

WITH STANDARD EXTENSION Z-ARM

TO ACHIEVE FLOOR

TO LIGHT DISTANCE

OF

“A”

IF DROP CEILING

IS

“B”

MOUNT BOTTOM OF

TRANSFORMER

BRACKET

FROM FLOOR AT

“C”

GAP BETWEEN
DROP CEILING

AND CEILING ARM

“D”

MINIMUM 2”

REQUIRED

75 INCHES

(1905 mm

)

8.0 FEET

(

2440 mm

)

99.8 INCHES

(2535 mm

)

6.5”

(165mm)

86 INCHES

(

2184 mm

)

9.0 FEET

(

2740 mm

)

109.8 INCHES

(

2789 mm

)

4.5”

(114mm)

______________________________________________________________________________

CEILING LIGHT MOUNTING INSTRUCTIONS

WITH EXTRA-HIGH EXTENSION Z-ARM

TO ACHIEVE

FLOOR TO LIGHT

DISTANCE OF

“A”

IF DROP CEILING

IS

“B”

MOUNT BOTTOM OF

TRANSFORMER

BRACKET

FROM FLOOR AT

“C”

GAP BETWEEN DROP

CEILING AND

CEILING ARM

“D”

MINIMUM 2”

REQUIRED

72 INCHES

(

1828 mm

)

10 FEET

(

2280 mm

)

123.8 INCHES

(

3145 mm

)

2.5”

(63mm)

85 INCHES

(

2032 mm

)

11 FEET

(

3350 mm

)

133.8 INCHES

(

3399 mm

)

4.5”

(114mm)