Conditions for the place of installation
Ambient temperature:
5°C to 40°C
If the ambient temperature drops below or
exceeds these values the electronic and
mechanical components can become damaged
and performance data are changed to an
unknown extent.
Atmospheric humidity:
Maximmum relative humidity 80% at temperatures up
to 31°C,
decreasing linearly down to 50% relative humidity at
40°C.
At higher atmospheric humidity the electronic
and mechanical components can become
damaged and performance data are changed
to an unknown extent.
Installation height:
max. 2000 m above sea level
Installation
Mount the RM200 only on a firm, stable laboratory
bench since otherwise vibrations can be transmitted.
Electrical connection
Voltage and frequency for the RM200 are given
on the type plate.
Ensure that these values correspond to the
available power supply system.
Connect the RM200 to the power supply system
using the supplied connection cable.
Protection by external fusing is to be used when
connecting the mains cable to the power supply,
according to the regulations at the place of
installation.
The voltage supply to the RM200 must be
provided with a fault-current circuit breaker (FI
switch).
If the values on the type plate are not observed this
can result in damage to the electrical and mechanical
components.