Conditions for the place of installation
Ambient temperature:
5°C to 40°C

If the ambient temperature drops below or

exceeds these values the electronic and

mechanical components can become damaged

and performance data are changed to an

unknown extent.

Atmospheric humidity:
Maximmum relative humidity 80% at temperatures up

to 31°C,
decreasing linearly down to 50% relative humidity at

40°C.

At higher atmospheric humidity the electronic

and mechanical components can become

damaged and performance data are changed

to an unknown extent.

Installation height:

max. 2000 m above sea level

Installation

Mount the RM200 only on a firm, stable laboratory

bench since otherwise vibrations can be transmitted.

Electrical connection

Voltage and frequency for the RM200 are given

on the type plate.

Ensure that these values correspond to the

available power supply system.

Connect the RM200 to the power supply system

using the supplied connection cable.

Protection by external fusing is to be used when

connecting the mains cable to the power supply,

according to the regulations at the place of

installation.

The voltage supply to the RM200 must be

provided with a fault-current circuit breaker (FI

switch).

If the values on the type plate are not observed this

can result in damage to the electrical and mechanical

components.