NTi Audio Acoustilyzer AL1 User Manual
Page 78

78
79
Firmware Update
NTi Audio provides free updates of the test instrument firmware and the
MiniLINK PC software on the NTi Audio website www.nti-audio.com.
To check out any available online updates, follow these instructions:
• Connect your PC to the web & start the MiniLINK PC software
• Connect the instrument with the USB cable to the PC
• Select the MiniLINK menu “Help → Look for Updates“
• Follow the download instructions
Alternatively you can check for any new firmware and software releases on
the web site
• Select the MiniLINK menu “Tools → Update Firmware“
Select the new firmware version in the displayed screen below:
for Minilyzer ML1:
ML1dA_V3.xx.bin
for Acoustilyzer:
AL1dA_A1.xx.bin
applicable firmware for
the connected device
After confirming the selected device firmware, the test instrument firmware
is updated automatically. In case of any problems during the firmware
upgrade, such as power failure, etc., the upgrade procedure can be
repeated anytime.
Note: The firmware update deletes the internal device memory.