New Era NE-1600 Multi-Phaser Programmable Syringe User Manual
Page 44
New Era Pump Systems Inc.
WWW.SYRINGEPUMP.COM
Model NE-1600/NE-1800 Multi-Phaser™
Publication #1200-02
39
8/1/14
OF
= Trigger off (disabled)
TTL I/O DIRECTIONAL CONTROL INPUT SETUP
DIN [ 0 | 1 ]
Set/query directional control input setup
Settings => 0 = Falling edge: Infuse, Rising edge Withdraw
Same as "dr:rE" setting from the keypad.
Use this setting with the CBL-TTL-1, reciprocating pump cable, to create a 2 pump
continuous infusion system
1 = Falling edge: Withdraw, Rising edge Infuse
Same as "dr:dU" setting from the keypad.
Use this setting with the CBL-TTL-1, reciprocating pump cable, to create a 2 pump
dual pumping system
PUMP MOTOR OPERATING TTL OUTPUT CONFIGURATION
ROM [
Set/query Pump Motor Operating TTL output configuration (TTL pin 7)
Settings => 0 = Output is logic high only when the pump motor is operating (pumping).
1 = Output is logic high when the pump motor is operating (pumping) or when the
Pumping Program is executing a pause timer.
SET KEYPAD LOCKOUT
LOC [P] [
Set/query keypad lockout mode. Set keypad lockout disables changing any settings from the keypad
unless the "Lockout Disable Key" is inserted.
P [
Set/Query Program Entry Mode Lockout. Set Program Entry Mode Lockout prevents inexperienced
users from entering “Program Entry Mode” from the keypad. When enabled, only the Phase 1 ‘Rate’,
‘Volume’ and Pumping Direction can be changed. Cannot be enabled when the Pumping Program is
currently programmed with a multiple Phase Program.
SET KEYPAD AND NOTIFICATION BEEP
BP [
Set/query keypad and notification beep mode. Set beep enables or disables beep mode.
9.4.4
General Control and Status Commands
TTL I/O OUTPUT SETTING
OUT
Sets TTL level on user definable output pin on the ‘TTL I/O’ connector.
Query is not applicable.
TTL INPUT QUERY
IN
Queries TTL level of pin on ‘TTL I/O’ connector. Set is not applicable.
Response:
BUZZER
BUZ [ 0 | { 1 [ < n > ] } ]
Sets / queries buzzer
Set:
0 = Turn buzzer off;
1 = Turn buzzer on
if
If
otherwise buzzer beeps
if