8/1/14
RAT I
Changes the pumping rate only if the pump is infusing. If the pumping direction is set to withdraw,
the rate change will be ignored.
VOLUME TO BE DISPENSED AND SET VOLUME UNITS
VOL [
Set/query volume to be dispensed:
Applicable only with “Rate Data Functions”. Can only be set when the Pumping Program is not
operating. The volume units are set according to the current syringe diameter setting. Do not send
the volume units when setting the volume.
Example: VOL 12.45
Sets the current phase “Volume to be Dispensed” to 12.45.
Set volume units:
Overrides the default volume units set when the syringe diameter is set.
Example: VOL UL
Sets all volume units to
µL.
Query response:
PUMPING DIRECTION
DIR [ INF | WDR | REV | STK ]
Set/query pumping direction
INF
= Infuse
WDR
= Withdraw
REV
= Reverse pumping direction
STK
= “Sticky Direction” (See “Sticky Direction”, sec: .5.8.1)
Applicable with all Program Phase functions. Cannot be set when the Pumping Program is operating
and the “Volume to be Dispensed” is non-zero.
The pumping direction cannot be changed if an alarm condition exists.
Query response: { INF | WDR | STK}
When set to “Sticky Direction”, query will respond with the current pumping direction (INF or WDR)
while pumping.
9.4.2
Pump Operational Commands
START PUMPING PROGRAM
RUN [
[E [
Starts the Pumping Program operation.
If the Pumping Program was paused, then the Pumping Program resumes at the point where it was
stopped. Otherwise, the Pumping Program starts from Phase 1.
Sub-Programs: If a Phase number is specified (
the specified Phase number. By programming sub-programs in different sections of the Pumping
Program memory, this command can be used to individually execute different sub-programs.
The pump cannot be started if an alarm condition exists.
E [
Trigger a Pumping Program Event while the Pumping Program is active.
Triggers a pre-defined event defined with the Pumping Program’s Event function, causing an
immediate jump to the Pumping Program Phase defined by the event function.