7 example 7: sub-programs – New Era NE-1600 Multi-Phaser Programmable Syringe User Manual
Page 34
New Era Pump Systems Inc.
WWW.SYRINGEPUMP.COM
Model NE-1600/NE-1800 Multi-Phaser™
Publication #1200-02
29
8/1/14
Phase
Function
9
LP:05
Phase
Function
10
LP:12
Phase
Function
11
JP:01
8.4.7
Example 7: Sub-Programs
This example shows some of the flexibilities provided by the Program Selection functions. The Pumping
Program starts by refilling the syringe with 50 mL at a fast pumping rate (Phase 1), then the Pumping Program
pauses for user sub-program selection (Phase 3). Then performs 5 dispenses of 10 mL at the selected rates,
then refills the syringe again and pauses for the next user sub-program selection.
The user is given the option of choosing one of three defined sub-programs.
1: Dispense 10 mL at 100 mL/hr (Phase 4)
2: Dispense 10 mL at 500 mL/hr (Phase 7)
3: Dispense 10 mL at 750 mL/hr (Phase 10)
After selecting the sub-program and pressing 'Start', the Pumping program continues execution at the selected
sub-program. After the 10 mL dispense, each sub-program jumps or continues with the loop counter function
(Phase 12). The first 4 loops continues Program Execution with the next user sub-program selection. After
the 5th loop, the program continues with Phase 13, which jumps back to the syringe refill function and starts
the whole program over.
Phase
Function
Rate
Volume
Direction
1
RATE
1500.0 mL/hr 50 mL
Withdraw
Phase
Function
2
LP:ST
Phase
Function
3
PR:IN
Phase
Function
4
PR:01
Phase
Function
Rate
Volume
Direction
5
RATE
100.0 mL/hr
10 mL
Infuse
Phase
Function
6
JP:12
Phase
Function
7
PR:02
Phase
Function
Rate
Volume
Direction
8
RATE
500.0 mL/hr
10 mL
Infuse
Phase
Function
9
JP:12
Phase
Function
10
PR:03
Phase
Function
Rate
Volume
Direction
11
RATE
750.0 mL/hr
10 mL
Infuse
Phase
Function
12
LP:05
Phase
Function
13
JP:01