14 error and alarm messages, 15 status messages – New Era NE-1600 Multi-Phaser Programmable Syringe User Manual
Page 16
New Era Pump Systems Inc.
WWW.SYRINGEPUMP.COM
Model NE-1600/NE-1800 Multi-Phaser™
Publication #1200-02
11
8/1/14
5.13.3
Default Program Pre-Load
Pressing the 'Volume'/'Program Function' key while turning on power to the pump will display the Default
Program Menu and the display will show:
. Currently, the only default program is the
Reciprocating Pump program. When selected, the following program will be pre-set, plus the RS-232
communications mode will be set to Reciprocating Mode. The default program can be modified as needed.
The following default program will be loaded.
The Reciprocating Pump Program sets the pump for use with a second pump, plus cable Part# CBL-DUAL, to
create a continuous infusion system.
Phase
Function
Rate
Volume
Direction
1
RATE
500 mL/hr
10.0 mL
Withdraw
Phase
Function
Rate
Volume
Direction
2
FILL
0.0 mL/hr
-----------
------------
Phase
Function
3
JP:01
5.13.4
Program Entry Mode Lockout
Pressing the ‘Diameter’ key while turning on power to the pump will enter special parameter setup. The
following will be displayed:
n
, currently, the only parameter. The “Program Entry Mode Lockout”,
when enabled, prevents inexperienced users from entering “Program Entry Mode” from the keypad. Mode
Disabled: ‘n’ = 0 (default). Mode Enabled: ‘n’ = 1. When enabled, only Phase 1 ‘Rate’, ‘Volume’ and
Pumping Direction can be changed. Cannot be enabled when the Pumping Program is currently programmed
with a multiple Phase Program.
5.14 Error and Alarm Messages
If the value entered is beyond the pumps capabilities or is invalid, or an operational problem occurred, one of
the following error or alarm messages will be displayed:
Pump motor stalled alarm.
Value entered is ‘Out Of Range’ of the pump’s operational limits.
nn
An out of range error occurred at Pumping Program Phase number ‘nn’, or the value just
entered is out of range. Check the pumping parameters and syringe diameter.
nn
A Pumping Program error was encountered at Phase number ‘nn’. The indicated Phase is
invalid in the context of the entire Pumping Program.
Key pressed is not currently applicable.
A communications time-out alarm occurred with an attached computer while operating in
the “Safe Communications Mode”. This most likely indicates that the RS-232 cable was
detached or the communication program on the computer has ended without turning off
“Safe Communications Mode”.
n
An error was detected during power up, where ‘n’ indicates the error. If n=1, then the
values stored in the pump’s non-volatile memory were invalid and were reset. If n=2,
then the non-volatile memory may need to be replaced.
Pump settings are locked out from the keypad. The lock out key is needed to change
settings. Lock out can also be reset with the reset function.
5.15 Status Messages
Indicates pumping rate units change mode.