New Era NE-1600 Multi-Phaser Programmable Syringe User Manual

Page 41

background image

New Era Pump Systems Inc.

WWW.SYRINGEPUMP.COM

Model NE-1600/NE-1800 Multi-Phaser™

Publication #1200-02

36

8/1/14

not operating

For a more detailed description of Program commands, see sec. 8.3, Program Function Descriptions.

=>

Rate Data Functions

When a Phase’s function is set to a “Rate Data Function”, use the ‘RAT’, ‘VOL’, and ‘DIR’
commands to setup the pumping parameters.

RAT

Pumping rate. ‘RATE’

FIL

Fill syringe to dispensed volume. ‘FILL’

INC

Increment rate. ‘INCR’

DEC

Decrement rate. ‘DECR’

Non-Rate Control Functions

STP

Stop pump. ‘STOP’

PAS

Pauses pumping for ‘nn’ seconds. ‘PS:nn’

PAS

Pauses pumping for 'n.n' seconds. 'PS:n.n'

PRI

Sub-Program Selection Input. 'Pr:In'

PRL

Sub-Program Start Label definition. 'Pr:nn'

LPS

Loop starting Phase. ‘LP:ST’

LPE

Loop end Phase. ‘LP:EN’

LOP

Loop to previous loop start ‘nn’ times. ‘LP:nn’

JMP

Jump to Program Phase. ‘JP:nn’

IF

If Program input TTL pin low, jump to Phase. ‘IF:nn’

EVN

Set event trigger trap. ‘EV:nn’

EVS

Set event square wave trigger trap. 'ES:nn'

EVR

Event trigger reset. ‘EV:RS’

CLD

Clear total dispense volume. ‘CLR.D’

TRG

Override Operational Trigger Default Mode, set to , ‘tr:aa’

BEP

Sound short beep. ‘BEEP’

OUT

Set programmable output pin. ‘OUT.n’

PUMPING RATE

RAT [ C | I ] [ [ ] ]

Set/query pumping rate.

=> UM

=

µL/min

MM

= mL/min

UH

=

µL/hr

MH

= mL/hr

Applicable only with “Rate Data Functions”.

When setting the pumping rate, if the current Phase’s function is not ‘RATE’, then is not
applicable.

While pumping, the pumping rate can only be set if the current Phase function is ‘RATE’ and the next
Program Phase’s function to be executed is not ‘INCR’ or ‘DECR’. Also, while pumping, units> cannot be set.

The new pumping rate will only be stored in non-volatile memory if the Pumping Program is not
operating.

When the pumping rate is queried while pumping, the response will be the current pumping rate and
units. Otherwise, the response will be the rate setting and units, if applicable. With the ‘INCR’ and
‘DECR’ functions, these two responses are not the same.

RAT C [ ]

Allows the Pumping Program to continue after the pumping rate is changed. The RAT C command
will not clear a Paused Pumping state after the rate is changed when the Pumping Program is paused.
Normally, the Pumping Program would re-start from Phase 1 after the rate is changed while paused.