Overview, 4 packing list, Features – New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual
Page 7
New Era Pump Systems Inc.
www.SyringePump.com
Model NE-500/NE-501 Multi-Phaser™
Publication #1200-02
2
09/18/14
option of New Era Pump Systems Inc., a defective unit will be either repaired or replaced.
This warranty does not cover damage by any cause including, but not limited to, any malfunction, defect or
failure caused by or resulting from unauthorized service or parts, improper maintenance, operation
contrary to furnished instructions, shipping or transit accidents, modifications or repair by the user, harsh
environments, misuse, neglect, abuse, accident, incorrect line voltage, fire, flood, other natural disasters, or
normal wear and tear. Changes or modifications not approved by New Era Pump Systems Inc. could void
the warranty. Wearable parts, such as drive nuts, are not covered by the warranty.
The foregoing is in lieu of all other expressed warranties and New Era Pump Systems Inc. does not assume
or authorize any party to assume for it any other obligation or liability.
1.4 Packing List
The OEM Starter Kit for the NE-500 / NE-501 Multi-Phaser™ Programmable Syringe Pump includes the
following items:
•
One of the following external unregulated power supply adapters:
Input: One of: 120V AC 60 Hz, 220V AC 50 Hz, 240V AC 50 HZ,
or other custom specified power supply
Output: 12V DC @ 800 mA (or compatible regulated power supply)
•
Primary Computer RS-232 interface cable: CBL-PC-PUMP-7
•
This User’s Manual
2. Overview
The model NE-500/NE-501 is a general purpose single syringe pump capable of infusion and withdrawal.
It is controlled from a microcontroller based system which drives a step motor, allowing a large range of
pumping rates calibrated to the inside diameter of the loaded syringe. The syringe is driven from a drive-
screw and drive-nut mechanism.
Features:
♦ Infusion and withdrawal pumping of syringes up
to 60 mL. 140 mL partially filled.
♦ Pumping rates from 0.73 µL/hr with a 1 mL
syringe to 2120 mL/hr with a 60 mL syringe.
♦ NE-501 only: Stall detection: Automatically
stops pump when pumping is impeded
♦ Infusion and withdrawal volumes separately
accumulated.
♦ Programmable dispense volumes.
♦ Programmable Phases allowing complex
pumping applications and interaction with
external devices.
♦ Program Sub-Programs, selectable by the user.
♦ Non-volatile memory of all operating
parameters and Pumping Program
♦ RS-232 bi-directional control from a computer.
♦ Built-in pump network driver. Pump network
supports up to 100 pumps and other devices.
♦ Two modes of RS-232 control, Basic and Safe.
Safe mode provides communication error
detection, loss of communication detection, and
automatic transmitting of alarm conditions.
♦ TTL logic I/O with firmware filtered control
inputs to eliminated glitches and ringing on the
control inputs.
♦ Configurable and programmable TTL
operational trigger for flexible logic control.
♦ Power Failure Mode: Restarts the Pumping
Program after a power interruption.
♦ Audible Alarm
♦ Many more features!