New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual
Page 25

New Era Pump Systems Inc.
www.SyringePump.com
Model NE-500/NE-501 Multi-Phaser™
Publication #1200-02
20
09/18/14
Phase
Function
4
LOP 50
Phase
Function
5
LPS
Phase
Function
Rate
Volume
Direction
6
DEC
1.0
0.1 mL
Infuse
Phase
Function
7
L0P 99
Phase
Function
Rate
Volume
Direction
8
DEC
1.0
0.1 mL
Infuse
Phase
Function
9
LPS
Phase
Function
Rate
Volume
Direction
10
INC
1.0
0.1 mL
Infuse
Phase
Function
11
LOP 50
Phase
Function
12
JMP 02
7.4.4
Example 4: Complex Dispenses with External Synchronization
0
0
750
W
it
h
d
ra
w
(
m
l/
h
r)
I
n
fu
s
e
Start
Trigger
Start
Trigger
:60
:60
:60
Start
Trigger
Start over
Syringe
Refill
A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate during
the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the pump waits for
a start trigger before starting the next dispense.
The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set of
dispenses, the syringe is refilled by the amount infused, 17.25 mL. Then the buzzer beeps, to alert the operator
to the start of the first set of dispenses. The process is then repeated.
Phase
Function
Rate
Volume
Direction
1
RAT
750.0 mL/hr
0.5 mL
Infuse
Phase
Function
Rate
Volume
Direction
2
RAT
300.0 mL/hr
1.5 mL
Infuse
Phase
Function
3
BEP
Phase
Function
4
PAS 00