Appendix, 1 rs-232 command summary – New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual
Page 46
New Era Pump Systems Inc.
www.SyringePump.com
Model NE-500/NE-501 Multi-Phaser™
Publication #1200-02
41
09/18/14
10. Appendix
10.1 RS-232 Command Summary
DIA [ < float > ]
Syringe inside diameter
PHN [ < phase data > ]
Program Phase number
FUN [ < phase function > ]
Program Phase function
< phase function > =>
RAT
Pumping rate.
FIL
Fill syringe to volume dispensed.
INC
Increment rate.
DEC
Decrement rate.
STP
Stop pump.
PAS
Pauses pumping for ‘nn’ seconds, if 0, wait for start trigger.
PAS [n.n]
Pauses pumping for 'n.n' seconds.
JMP
Jump to Program Phase.
PRI
Sub-Program Selection Input from expansion connector
PRL
Sub-Program Start Label definition.
LPS
Loop starting Phase.
LOP
Loop to previous loop start ‘nn’ times.
LPE
Loop end Phase.
IF
If Program input low, jump to Program Phase.
EVN
Set event trigger.
EVS
Set event square wave trigger.
EPL
Pause program, wait for low edge or level on expansion pin
EPE
Pause program, wait for falling edge on expansion pin
EVE
Re-direct event trigger to expansion pin
EVR
Event trigger reset.
CLD
Clear total dispense volume.
OUT { 0 | 1 }
Set programmable output pin.
OE0
Set expansion pin low
OE1
Set expansion pin high
TRG
Override Operational Trigger default setting
BEP
Sound short beep.
RAT {C | I ] [
VOL [
Volume to be Dispensed, or set Volume units
DIR [ INF | WDR | REV | STK ]
Pumping direction
RUN [
Starts the Pumping Program
[E [
Pumping Program event trigger
PUR
Start purge
STP
Stop/pauses the Pumping Program
DIS
Query volume dispensed
CLD { INF | WDR }
Clear volume dispensed
SAF [
Safe communications mode
LN [ 0 | 1]
Low motor noise mode
AL [ 0 | 1]
Alarm buzzer mode
PF [ 0 | 1]
Power failure auto-restart mode
TRG [ FT | FH | F2 | LE | ST | T2 | SP | P2 | RL | RH | SL | SH | OF ]
Operational trigger default mode
DIN [ 0 | 1]
Directional input control mode
ROM [0 | 1]
Pump Motor Operating TTL output mode
LOC [ P ] [ 0 | 1]]
Keypad lockout mode or Program Entry Mode lockout
BP [ 0 | 1]
Key and Notification beep mode
OUT [ E
Set TTL output level
IN { E
Query TTL input level
BUZ [ 0 | { 1 [ < n > ] } ]
Buzzer control
VER
Query firmware version
System Commands: Valid regardless of current network address
*ADR [
Network address and baud rate