7.4 Pumping Program Examples
7.4.1
Example 1: 2 Step Rate
ml Dispensed
0 5
30
0
2.5
500
m
l/
h
r
R
a
te
Infuse 5.0 mL at 500 mL/hr, and then infuse 25.0 mL at 2.5 mL/hr. Then stop the pump.
Phase
Function
Rate
Volume
Direction
1
RAT
500 mL/hr
5.0 mL
Infuse
Phase
Function
Rate
Volume
Direction
2
RAT
2.5 mL/hr
25.0 mL
Infuse
Phase
Function
3
STP
To send this program to the pump, send the following sequence of commands:
PHN 1
FUN RAT
RAT 500 MH
VOL 5.0
DIR INF
PHN 2
FUN RAT
RAT 2.5 MH
VOL 25.0
DIR INF
PHN 3
FUN STP
7.4.2
Example 2: Repeated Dispenses with Suck Back
m
l/
h
r
R
a
te
Time
0
0
750
5:00
:30
Dispense 2.0 mL with a 5 minute pause between dispenses. In addition, after each dispense, a volume of 0.25
mL is sucked back to prevent dripping. Also, 30 seconds before the end of the pause interval, a beep is
sounded to alert the operator to prepare for the next dispense.
Starting with the second dispense, 0.25 is added to the volume dispensed to compensate for the sucked back
volume of the previous dispense. By changing the last Phase to a [FUN LOP nn] function, the total number of
dispenses can be set.