New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual
Page 27
![background image](/manuals/569076/27/background.png)
New Era Pump Systems Inc.
www.SyringePump.com
Model NE-500/NE-501 Multi-Phaser™
Publication #1200-02
22
09/18/14
When the low pressure trap is triggered, the pump sets the high pressure trap (PHASE 07) and begins to
increment the flow rate. The flow rate is incremented in 1.0 mL/hr steps with every 0.25 mL dispensed
(PHASE 08-10). If the high pressure trap hasn’t as yet been triggered, the flow rate will max out at 25.0 mL/hr
while waiting for the high pressure trap (PHASE 11). When the high pressure point is reached, the pump
immediately will drop down to 10.0 mL/hr (PHASE 02), and once again wait for the low pressure point.
Phase
Function
1
OUT 0
Phase
Function
Rate
Volume
Direction
2
RAT
10.0 mL/hr
0.005
Infuse
Phase
Function
3
EVN 05
Phase
Function
Rate
Volume
Direction
4
RAT
10.0 mL/hr
0.0 mL (off)
Infuse
Phase
Function
5
OUT 1
Phase
Function
Rate
Volume
Direction
6
RAT
10.0 mL/hr
0.005
Infuse
Phase
Function
7
EVN 01
Phase
Function
8
LPS
Phase
Function
Rate
Volume
Direction
9
INC
1.0
0.25 mL
Infuse
Phase
Function
10
L0P 14
Phase
Function
Rate
Volume
Direction
11
RAT
25.0 mL/hr
0.0 mL (off)
Infuse
7.4.6
Example 6: Automated Dispensing with Synchronization
'Power Failure
Mode' Pumping
Program Restart
at Power Up
Syringe
Refill
Event Trigger
from
Syringe Filled Sensor:
Jump to Phase 3
5:00:00
5:00:00
5:00:00
5:00:00
5:00:00
Restart Pumping Program
The following is an automated dispensing Program for a NE-500/501 syringe pump equipped with a ‘Syringe
Filled Sensor’ attached to the TTL I/O connector, and a valve system to refill the syringe from a reservoir. It is
also assumed that the ‘Power Failure’ mode is enabled.
After a power fail restart, the pusher block is in an unknown position, making it impossible for an automated
dispensing system to regain synchronization. With a Syringe Filled Sensor, the following Pumping Program
will automatically synchronize the dispensing system, and then continue with the normal dispense.