0 run the test, 0 turn on hydraulic power, 0 perform crack size check (csc) – MTS Multipurpose Simulated User Manual
Page 91

Load P Parameters
Description
Parameter
Specify the crack size value to use in the load
calculation.
Crack Size
Specify the stress intensity (K) value to use
in the load calculation.
Stress Intensity K
The following table lists the stress intensity K parameters.
Stress Intensity K Parameters
Description
Parameter
Specify the crack size value to use in the
stress intensity (K) calculation.
Crack Size
Specify the load value to use in the stress
intensity (K) calculation.
Load
5.6.3.0 Run the Test
5.6.3.1.0 Turn On Hydraulic Power
Use the control panel to apply hydraulic power.
1. If the Interlock indicator is on, click Reset to release system interlocks.
If the interlock remains on, use the message logs to identify and correct the cause of the interlock.
2. Click Low Power.
This action sets the hydraulic power unit to high pressure and the hydraulic service manifold to low pressure.
3. Click High Power.
This action sets the hydraulic service manifold to high pressure.
5.6.3.2.0 Perform Crack Size Check (CSC)
This activity applies a load to the specimen and verifies the crack size and the modulus for tests that use a
COD clip gage to measure the crack opening. For description of the CSC parameters and their values, see
on page 92.
1. Click CSC on the Main menu.
2. Optionally, click Change Parameters. These parameters determine how to measure crack size, and how
to calculate crack size and modulus.
3. Click Measure Crack Size.
MTS TestSuite 91
5.0 Running Simulation Tests