0 jic parameters – MTS Multipurpose Simulated User Manual
Page 74

Precrack Parameters
Description
Parameter
Specify the desired crack size to achieve in
the precrack activity. When the specified
crack size is achieved, the activity stops.
Precrack Final Crack
Limit
Specify the cycle frequency of the command
signal for the precrack activity.
Precrack Frequency
Specify the ratio of the minimum to maximum
load applied to the specimen. The minimum
Precrack Load Ratio
load is determined by this value and the
maximum load is specified by you.
Specify the percentage of the measured load
range used as the lower limit for linear least
Precrack Lower Least
Squares Fit Percentage
squares fit compliance determination. This
parameter is not used for FCG DCPD tests.
Specify the percentage of the measured load
range used as the upper limit for linear least
Precrack Upper Least
Squares Fit Percentage
squares fit compliance determination. This
parameter is not used for FCG DCPD tests.
Specify the load level at which the voltage is
measured. The load is applied at this level
Precrack Measure Load
Level
to ensure the crack is open. This parameter
is used only for FCG DCPD tests.
Specify the number of cycles at which
precrack activity stops.
Precrack Cycle Limit
Specify the maximum stress intensity desired
at the end of precrack activity. The initial
Precrack Final Maximum
K
precrack Maximum K is 1.4 times the Final
Maximum K value.
5.5.2.4.0 JIC Parameters
In the JIC Parameters section of the main menu, you select COD (Crack Opening Displacement) or
Displacement as the control mode for the test, and then specify the parameters for the selected mode.
74 MTS TestSuite