5.3.2.7.0 Data Storage Parameters Properties
Data Storage Parameters
Description
Parameter
Specify the percentage of precrack activity
that must complete before data is saved to
disk.
Precrack Save Percent
Limit
Specify the cycle intervals written to disk
during the precrack activity.
Store Every Nth Pre-crack
Cycle
Specify the lower load limit for linear least
squares fit compliance used to determine
FCG Lower Least Squares
Fit Percentage
when to start saving data to disk. This
parameter is not used for FCG DCPD tests.
Specify the upper load limit for linear least
squares fit compliance used to determine
FCG Upper Least Squares
Fit Percentage
when to start saving data to disk. This
parameter is not used for FCG DCPD tests.
Specify the cycle intervals written to disk
during the test.
FCG Nth Cycle Save
5.3.2.8.0 Test Termination Parameters
Test Termination Parameters
Description
Parameter
Specify Yes to end the test when the COD
limit is reached.
COD Limit
Specify Yes to end the test when the load
limit is reached.
Load Limit
Specify Yes to end the test when the
displacement limit is reached.
Displacement Limit
5.3.2.9.0 Assign Precrack Values
Specify these parameters if the specimen was precracked on another load frame. When the precrack
parameters are specified here, you can then skip Precrack Specimen in the Main Menu window.