0 set the fgc dcpd test parameters, 0 test parameters overview, 0 fcg dcpd parameter values for simulation – MTS Multipurpose Simulated User Manual
Page 57

5. Reset the interlocks if needed and select the Low, then High, power buttons.
6. Click Run to run the test.
5.4.2.0 Set the FGC DCPD Test Parameters
5.4.2.1.0 Test Parameters Overview
After a new test run is added to the test, the Main Menu window appears. This window provides access to
all necessary test parameters. Click the parameter buttons in the Main Menu window to specify parameter
values for the test.
5.4.2.2.0 FCG DCPD Parameter Values for Simulation
The sample test includes a test run with test data that can be analyzed. If you want to reproduce the same
results in another test run, add another test run, and then specify the following values for the test parameters.
Other parameters not listed here should remain unchanged.
Precrack Parameters
Precrack Parameter Values
Value to Use for Simulation
Parameter
10 Hz
Precrack Frequency
25 mm
Precrack Final Crack Limit
1 kN
Precrack Measure Load Level
0.1 unitless
Precrack Load Ratio
0.1 kN/mm^1.5
Precrack Final Maximum K
1000000 count
Precrack Cycle Limit
Crack Growth Parameters
Delta K Parameter Values
Value to Use for Simulation
Parameter
True Sine
FCG Wave Shape
10 Hz
FCG Frequency
0.2 kN/mm^1.5
FCG Delta-K Initial
0.1 unitless
FCG Load Ratio
MTS TestSuite 57
5.0 Running Simulation Tests