0 precrack parameters – MTS Multipurpose Simulated User Manual
Page 88

Value to Use for Simulation
Parameter
1.35 mm
KIC COD Upper Limit
No
KIC Use Load Limit
200000 N
KIC Load Upper Limit
No
KIC Use Displacement Limit
50 mm
KIC Displacement Upper Limit
5.6.2.3.0 Precrack Parameters
The following table lists the precrack parameters and their required values.
Precrack Parameters
Description
Parameter
Specify the desired crack size to achieve in
the precrack activity. When the specified
crack size is achieved, the activity stops.
Precrack Final Crack
Limit
Specify the cycle frequency of the command
signal for the precrack activity.
Precrack Frequency
Specify the ratio of the minimum to maximum
load applied to the specimen. The minimum
Precrack Load Ratio
load is determined by this value and the
maximum load is specified by you.
Specify the percentage of the measured load
range used as the lower limit for linear least
Precrack Lower Least
Squares Fit Percentage
squares fit compliance determination. This
parameter is not used for FCG DCPD tests.
Specify the percentage of the measured load
range used as the upper limit for linear least
Precrack Upper Least
Squares Fit Percentage
squares fit compliance determination. This
parameter is not used for FCG DCPD tests.
Specify the load level at which the voltage is
measured. The load is applied at this level
Precrack Measure Load
Level
to ensure the crack is open. This parameter
is used only for FCG DCPD tests.
Specify the number of cycles at which
precrack activity stops.
Precrack Cycle Limit
88 MTS TestSuite