beautypg.com

2 instrument malfunctions, Troubleshooting, Instrument malfunctions – Leica Biosystems RM2235 User Manual

Page 68

background image

66

Instructions for Use V 1.8 RevC – 09/2013

Problem

Possible cause

Corrective action

4. Noises during sectioning

The knife "sings" when section-

ing hard specimens. The sec-

tions show scratches or chatter

marks.

• The sectioning speed is too fast.

• The clearance angle is too wide.

• Insufficient clamping of speci-

men and/or knife holder.

• Turn the handwheel at a slower

speed.

• Systematically decrease the

clearance angle until the opti-

mum adjustment is obtained.

• Check all screw and clamp con-

nections on the specimen holder

system and the knife holder. If

necessary, tighten the levers and

screws.

7.2 Instrument malfunctions

1. There is no further feed motion

and thus no sectioning.

• The front end position has been

reached.

• Run the specimen back by turn-

ing the coarse driving wheel.

2. High blade consumption

• Too great of a sectioning force

was applied.

• Adjust the sectioning speed and/

or section thickness when trim-

ming. Select a smaller section

thickness, rotate the handwheel

more slowly.

7. Troubleshooting